9184 sc HOCL Chlorine Hypochlorous Chlorine (HOCl) amperometric analyser

9184 sc HOCL Chlorine Hypochlorous Chlorine (HOCl) amperometric analyser
Codice articolo: LXV430.99.00001
Prezzo unitario Contatti
Spedizione entro 1 settimana

Low Minimum Detection Limit for Efficient Residual Chlorine Monitoring

The Hach 9184sc sensor has a minimum detection limit of 5 ppb or 0.005 mg/L hypochlorous acid in the form of HOCl.

Wide Measurement Range

The 9184sc sensor is particularly useful for drinking water distribution monitoring, chlorination applications, demineralizer systems, wastewater, or cooling water processes due to its 0 - 20 ppm measurement range.

Continuous Chlorine Measurement

Measurements are made continuously by the sensor and reported to the controller

Compatible with Hach's "Plug and Play" Digital Controllers

The 9184sc Analyser is designed to be used with all Hach SC digital controllers. Set up is fast and easy, and the system is compatible with the full line of Hach digital sensors.

Contenuto della confezione

9184sc Chlorine Sensor Panel, Membranes and Electrolyte and Manual. Does not include pH sensor.

Specifiche tecniche

Accuratezza: 2 % or ±10 ppb HOCl, whichever is greater at pH <8
  2 % or ±10 ppb TFC, whichever is greater at pH <7.5
Compensazione della temperatura: Automatic over sample temperature range
  Range di temperatura automatico
sul campione
Condizioni di stoccaggio: -20 °C - 60 °C
Contenuto della confezione : 9184sc Chlorine Sensor Panel, Membranes and Electrolyte and Manual. Does not include pH sensor.
Descrizione: 9184sc Hypochlorous Acid Chlorine Amperometric Sensor
Deviazione standard: 0.7 %
Flusso: 0.25 L/h (200 - 250 mL/min) auto--regulated by flow through cell
Garanzia: 24 mesi
Grado di protezione : IP66 / NEMA 4X
Ingresso del campione: 0.25" outer diameter
Interferenze: No interference from chloramines.
Chlorine dioxide and ozone interfere with measurement.
Intervallo di manutenzione: Measurement cell: 6 months for membrane and electrolyte for typical application
pH: 1 - 1.5 years for typical application
Intervallo di misura: 0 - 20 mg/L hypochlorous acid HOCl
  Continuous
Intervallo di pressione: 0.1 - 2 bar in flow cell
Intervallo di taratura: 2 mesi
Intervallo pH: 4 - 8 unità di acidificazione disponibile per pH
superiore a 8
Limite di rilevamento: 0.005 mg/L HOCl
Metodo di taratura: Zero Calibration: electrically (automatic) or with de-chlorinated water
Calibration: Comparison of lab method with process sample
pH Calibration: Single- or two-point using standards or comparison of lab method with process sample
Montaggio: Flat, vertical surface (panel, stand, etc.)
Parametro: Active Free Chlorine: Hypochlorous Acid Only
Peso: 6,5 kg 14,3 lbs
Principio di misura: Amperometric/Membrane (electrode, membrane, electrolyte)
Temperatura del campione: 2 - 45°C (35,6 - 113°F)
Temperatura di lavoro: 0 - 45 °C
Tempo di risposta: 90 % in less than 90 seconds
Umidità relativa: 0 - 90 % non-condensing
Uscita del campione: 0.5" inner diameter